Baxter Healthcare Corporation: Welch Allyn INF BAG, Thigh 1-Tube
Recall reason: Replacement bladder kit for the Welch Allyn two-piece reusable blood pressure cuff may contain a Large Adult size bladder rather than a Thigh size bladder in the replacement kits.. Recalled by Baxter Healthcare Corporation. Distribution: US Nationwide distribution in the states of California, Colorado, Ohio, and Utah.. Classification (Class II — potential health risk). Affected lots/dates: UDI- DI 00732094110845, Lot Numbers: 23-296, 23-318, 23-325, 23-353, 24-008, 24-036, 24-037, 24-138, 24-183, 25-037, 25-. Contact the manufacturer or the retailer where you purchased this device to arrange for a free correction, repair, or replacement.